Important Control Solution Information
Use only UltraTRAK control solutions with your meter.
Do not use the control solution beyond the expiry date or 1.5
months after rst opening. Write the opening date on the control
solution vial and discard the remaining solution after 1.5 months.
It is recommended that the control solution test be done at room
temperature (68°F-77°F / 20°C-25°C). Make sure your control
solution, meter, and test strips are at this specied temperature
range before testing.
Shake the vial before use, discard the rst drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
Store the control solution tightly closed at temperatures between
36°F and 86°F (2°C and 30°C). Do NOT freeze.
Do not use test strips beyond the expiry date. This may cause
inaccurate results.
Do not bend, cut, or alter a test strip in any way.
Keep the strip vial away from children since the cap and the test
strip may be a choking hazard. If swallowed, promptly see a doctor
for help.
For further information, please refer to the test strip package insert.