Error Messages
MESSAGE WHAT IT MEANS WHAT TO DO
Appears when the batteries
cannot provide enough
power for a test.
Replace the batteries immediately.
Appears when a used
test strip is inserted.
Repeat with a new test strip.
Problem in operation.
Repeat the test with a new
test strip.
If the meter still does not work,
please contact customer ser-
vices for assistance.
Appears when test strip is
removed while counting
down, or insufcient blood
volume.
Repeat the test with a new
test strip.
Appears when ambient
temperature is below
system operation range.
System operation range is
50
o
F to 104
o
F (10
o
C to 40
o
C).
Repeat the test after the
meter and test strip are in the
above temperature range.
Appears when ambient
temperature is above
system operation range.